Remote Commands in the Receiver Application
R&S
®
ESW
536User Manual 1177.6298.02 ─ 08
Parameters:
<Percentage> Statistical frequency percentage.
Range: 0 to 66
*RST: 0
Default unit: %
Example:
DISP:WIND:SGR:COL:UPP 95
Sets the start of the color map to 95%.
Manual operation: See "Start / Stop" on page 218
DISPlay[:WINDow<n>]:SGRam:COLor[:STYLe]
<ColorScheme>
DISPlay[:WINDow<n>]:SPECtrogram:COLor[:STYLe] <ColorScheme>
This command selects the color scheme.
Parameters:
<ColorScheme> HOT
Uses a color range from blue to red. Blue colors indicate low lev-
els, red colors indicate high ones.
COLD
Uses a color range from red to blue. Red colors indicate low lev-
els, blue colors indicate high ones.
RADar
Uses a color range from black over green to light turquoise with
shades of green in between.
GRAYscale
Shows the results in shades of gray.
*RST:
HOT
Example:
DISP:WIND:SPEC:COL GRAY
Changes the color scheme of the spectrogram to black and
white.
Manual operation: See "Hot/Cold/Radar/Grayscale" on page 219
15.7.3.6 Formats for Returned Values: ASCII Format and Binary Format
When trace data is retrieved using the TRAC:DATA or TRAC:IQ:DATA command, the
data is returned in the format defined using the FORMat[:DATA] on page 580. The
possible formats are described here.
●
ASCII Format (FORMat ASCII):
The data is stored as a list of comma-separated values (CSV) of the measured val-
ues in floating point format.
●
Binary Format (FORMat REAL,32):
The data is stored as binary data (Definite Length Block Data according to IEEE
488.2), each measurement value being formatted in 32-Bit IEEE 754 Floating-
Point-Format.
The schema of the result string is as follows:
Analysis