Remote Commands in the Receiver Application
R&S
®
ESW
511User Manual 1177.6298.02 ─ 08
15.6.6 Bandwidth and Filter Configuration
[SENSe:]BWIDth:IF........................................................................................................ 511
[SENSe:]BANDwidth:IF...................................................................................................511
[SENSe:]BWIDth[:RESolution]......................................................................................... 511
[SENSe:]BANDwidth[:RESolution][:VALue]........................................................................511
[SENSe:]BWIDth:SCPL...................................................................................................511
[SENSe:]BANDwidth:SCPL............................................................................................. 511
[SENSe:]BWIDth[:RESolution]:AUTO...............................................................................512
[SENSe:]BANDwidth[:RESolution]:AUTO..........................................................................512
[SENSe:]BWIDth[:RESolution]:TYPE................................................................................512
[SENSe:]BANDwidth[:RESolution]:TYPE.......................................................................... 512
[SENSe:]BWIDth:IF <Bandwidth>
[SENSe:]BANDwidth:IF <Bandwidth>
This command defines the resolution bandwidth for IF analysis.
Parameters:
<Bandwidth> <numeric value>
The available bandwidths depend on the span.
Default unit: Hz
Example: //Define resolution bandwidth for IF analysis
BAND:IF 10KHZ
Manual operation: See "IF Analysis" on page 181
[SENSe:]BWIDth[:RESolution] <Bandwidth>
[SENSe:]BANDwidth[:RESolution][:VALue] <Bandwidth>
This command defines the measurement (or resolution) bandwidth.
The available bandwidths depend on the selected filter type.
A change of the resolution bandwidth automatically turns the coupling to the frequency
off.
Parameters:
<Bandwidth> <numeric value>
Refer to the datasheet for available bandwidths.
Default unit: Hz
Example: //Select measurement bandwidth
BAND 3MHZ
Manual operation: See "Measurement Bandwidth" on page 101
[SENSe:]BWIDth:SCPL <State>
[SENSe:]BANDwidth:SCPL <State>
This command couples or decouples the IF span to the measurement bandwidth used
for the bargraph.
Configuration