Remote Commands in the Receiver Application
R&S
®
ESW
512User Manual 1177.6298.02 ─ 08
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON
Example: //Decouple span from the measurement bandwidth.
BAND:SCPL OFF
Manual operation: See "IF Analysis" on page 181
[SENSe:]BWIDth[:RESolution]:AUTO <State>
[SENSe:]BANDwidth[:RESolution]:AUTO <State>
This command couples or decouples the resolution bandwidth to the selected fre-
quency.
The resolution bandwidth is coupled to the frequency only if you have selected the
Quasipeak, CISPR Average or CISPR RMS detector.
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON
Example: //Decouple measurement bandwidth from frequency range
BAND:AUTO OFF
Manual operation: See "CISPR RBW Uncoupled" on page 182
[SENSe:]BWIDth[:RESolution]:TYPE <FilterType>
[SENSe:]BANDwidth[:RESolution]:TYPE <FilterType>
This command selects the resolution filter type.
When you change the filter type, the command selects the next larger filter bandwidth if
the same bandwidth is unavailable for that filter.
Parameters:
<FilterType> CFILter
Channel filters
Available in the spectrum application.
NORMal
Gaussian filters
P5
5-pole filters
Available for FFT sweeps in the spectrum application.
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB)
Return value for query is always PULS.
MIL
MIL Std (6 dB)