Remote Commands in the Receiver Application
R&S
®
ESW
513User Manual 1177.6298.02 ─ 08
*RST: NORMal
Example:
BAND:TYPE NORM
Manual operation: See "Filter Type" on page 101
15.6.7 Trigger Configuration
[SENSe:]SWEep:EGATe:HOLDoff....................................................................................513
[SENSe:]SWEep:EGATe:LENGth.....................................................................................513
[SENSe:]SWEep:EGATe:POLarity....................................................................................513
[SENSe:]SWEep:EGATe:SOURce................................................................................... 514
[SENSe:]SWEep:EGATe................................................................................................. 514
[SENSe:]SWEep:EGATe:TYPE........................................................................................515
TRIGger<tp>[:SEQuence]:HOLDoff[:TIME].......................................................................515
TRIGger<tp>[:SEQuence]:LEVel[:EXTernal]......................................................................515
TRIGger<tp>[:SEQuence]:SLOPe....................................................................................516
TRIGger<tp>[:SEQuence]:SOURce................................................................................. 516
[SENSe:]SWEep:EGATe:HOLDoff <DelayTime>
This command defines the delay time between the gate signal and the continuation of
the measurement.
Parameters:
<DelayTime> Range: 0 s to 30 s
*RST: 0 s
Default unit: S
Example:
SWE:EGAT:HOLD 100us
Manual operation: See "Gate Delay" on page 188
[SENSe:]SWEep:EGATe:LENGth <GateLength>
This command defines the gate length.
Parameters:
<GateLength> Range: 125 ns to 30 s
*RST: 400μs
Default unit: S
Example:
SWE:EGAT:LENG 10ms
Manual operation: See "Gate Length" on page 189
[SENSe:]SWEep:EGATe:POLarity <Polarity>
This command selects the polarity of an external gate signal.
The setting applies both to the edge of an edge-triggered signal and the level of a
level-triggered signal.