MMEMory:SELect:CHANnel[:ITEM]:TRANsducer:ALL <State>
MMEMory:SELect[:ITEM]:TRANsducer:ALL <arg0>
This command includes or excludes transducer factors when storing or loading a con-
figuration file.
Parameters:
<arg0> ON | OFF | 1 | 0
*RST: 0
Example:
MMEM:SEL:TRAN:ALL ON
Manual operation: See "Items:" on page 266
15.8.3
Instrument Setting Management
MMEMory:CLEar:ALL.....................................................................................................589
MMEMory:CLEar:STATe................................................................................................. 589
MMEMory:LOAD:AUTO..................................................................................................589
MMEMory:LOAD:STATe..................................................................................................590
MMEMory:STORe<1|2>:STATe....................................................................................... 591
MMEMory:STORe<1|2>:STATe:NEXT..............................................................................592
MMEMory:STORe<1|2>:TYPE........................................................................................ 592
SYSTem:PRESet........................................................................................................... 593
SYSTem:PRESet:CHANnel[:EXEC]................................................................................. 593
MMEMory:CLEar:ALL
This command deletes all <instrument> configuration files in the current directory.
You can select the directory with MMEMory:CDIRectory on page 582.
Example:
MMEM:CLE:ALL
MMEMory:CLEar:STATe <arg0>, <arg1>
This command deletes an instrument configuration file.
Parameters:
<arg0>
<arg1> String containing the path and name of the file to delete.
The string may or may not contain the file's extension.
Example:
MMEM:CLE:STAT 1,'TEST'
MMEMory:LOAD:AUTO <arg0>, <arg1>
This command restores an <instrument> configuration and defines that configuration
as the default state.
The default state is restored after a preset (*RST) or after you turn on the R&S ESW.