EasyManua.ls Logo

Rohde & Schwarz ESW Series - Instrument Setting Management

Rohde & Schwarz ESW Series
759 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands in the Receiver Application
R&S
®
ESW
589User Manual 1177.6298.02 ─ 08
MMEMory:SELect:CHANnel[:ITEM]:TRANsducer:ALL <State>
MMEMory:SELect[:ITEM]:TRANsducer:ALL <arg0>
This command includes or excludes transducer factors when storing or loading a con-
figuration file.
Parameters:
<arg0> ON | OFF | 1 | 0
*RST: 0
Example:
MMEM:SEL:TRAN:ALL ON
Manual operation: See "Items:" on page 266
15.8.3
Instrument Setting Management
MMEMory:CLEar:ALL.....................................................................................................589
MMEMory:CLEar:STATe................................................................................................. 589
MMEMory:LOAD:AUTO..................................................................................................589
MMEMory:LOAD:STATe..................................................................................................590
MMEMory:STORe<1|2>:STATe....................................................................................... 591
MMEMory:STORe<1|2>:STATe:NEXT..............................................................................592
MMEMory:STORe<1|2>:TYPE........................................................................................ 592
SYSTem:PRESet........................................................................................................... 593
SYSTem:PRESet:CHANnel[:EXEC]................................................................................. 593
MMEMory:CLEar:ALL
This command deletes all <instrument> configuration files in the current directory.
You can select the directory with MMEMory:CDIRectory on page 582.
Example:
MMEM:CLE:ALL
MMEMory:CLEar:STATe <arg0>, <arg1>
This command deletes an instrument configuration file.
Parameters:
<arg0>
<arg1> String containing the path and name of the file to delete.
The string may or may not contain the file's extension.
Example:
MMEM:CLE:STAT 1,'TEST'
MMEMory:LOAD:AUTO <arg0>, <arg1>
This command restores an <instrument> configuration and defines that configuration
as the default state.
The default state is restored after a preset (*RST) or after you turn on the R&S ESW.
Data Management

Table of Contents

Related product manuals