Remote Commands in the Receiver Application
R&S
®
ESW
472User Manual 1177.6298.02 ─ 08
15.5.9 Final Measurement (and Trace) Configuration
CALCulate<n>:FMEasurement[:AUTO]............................................................................ 472
[SENSe:]CORRection:TRANsducer:VIEW........................................................................ 472
[SENSe:]DETector<t>:FMEasurement..............................................................................473
[SENSe:]FMEasurement:AUTO.......................................................................................473
[SENSe:]FMEasurement:TIME........................................................................................ 473
DISPlay[:WINDow<n>]:TRACe<t>:MODE.........................................................................474
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]............................................ 475
CALCulate<n>:FMEasurement[:AUTO] <State>
This command turns a full automated test sequence on and off.
A full automated test sequence includes a scan, a peak search and a final measure-
ment.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Turn on automated test sequence.
CALC:FME ON
[SENSe:]CORRection:TRANsducer:VIEW <State>
This command turns the "Transducer" trace mode on and off.
Also possible with DISPlay[:WINDow<n>]:TRACe<t>:MODE on page 474.
Parameters:
<State> ON | 1
Turns on the transducer trace.
The transducer trace represents the correction values of all
active transducers over the currently selected frequency range.
When you turn on the transducer trace, all other traces are tem-
porarily removed.
OFF | 0
Turns off the transducer trace, and restores the original scan
traces.
*RST: 0
Example:
CORR:TRAN:VIEW ON
TRAC? TRACE1
Turns on the transducer trace and queries the correction values.
Manual operation: See "Trace Mode" on page 132