Remote Commands in the Receiver Application
R&S
®
ESW
474User Manual 1177.6298.02 ─ 08
Parameters:
<Time> <numeric value>
*RST: 1 s
Default unit: s
Example: //Define final measurement time
FME:TIME 1us
Manual operation: See "Measurement Time" on page 100
DISPlay[:WINDow<n>]:TRACe<t>:MODE
<DisplayMode>
This command selects the trace mode.
To turn on the transducer trace mode, use [SENSe:]CORRection:TRANsducer:
VIEW on page 472.
Suffix:
<n>
.
irrelevant
<t> Trace
Parameters:
<DisplayMode> AVERage
Draws a trace based on the average over several measure-
ments.
You can define the number of measurements considered in the
averaging process with [SENSe:]SWEep:COUNt.
Unavailable in Receiver mode.
BLANk
Removes the trace from the display.
MAXHold
Draws a trace based on the highest values that have been mea-
sured over several measurements.
You can define the number of measurements with [SENSe:
]SWEep:COUNt.
MINHold
Draws a trace based on the lowest values that have been mea-
sured over several measurements.
You can define the number of measurements with [SENSe:
]SWEep:COUNt.
TRD
Draws a trace that shows the correction values of all active
transducer factors in the currently selected frequency range.
See also [SENSe:]CORRection:TRANsducer:VIEW
on page 472.
VIEW
Freezes the trace.
Even if you continue the measurement, the trace remains as it
is.
Measurements and Result Displays