MMEMory:STORe<1|2>:STATe:NEXT
This command saves the current <instrument> configuration in a *.dfl file.
The file name depends on the one you have set with MMEMory:STORe<1|2>:STATe
on page 591. This command adds a consecutive number to the file name.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256 MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see Chapter 4.7, "Protecting Data Using the Secure User Mode",
on page 40.
Suffix:
<1|2>
.
irrelevant
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current <instrument> settings in the file Save.dfl.
MMEM:STOR:STAT:NEXT
Saves the current <instrument> settings in the file
Save_001.dfl
MMEM:STOR:STAT:NEXT
Saves the current <instrument> settings in the file
Save_002.dfl
Manual operation: See "Save File" on page 266
MMEMory:STORe<1|2>:TYPE <arg0>
This command defines whether the data from the entire <instrument> or only from the
current channel is stored with the subsequent MMEM:STOR... command.
Suffix:
<1|2>
.
irrelevant
Parameters:
<arg0> INSTrument | CHANnel
INSTrument
Stores data from the entire <instrument>.
CHANnel
Stores data from an individual channel.
*RST: INST
Example:
INST:SEL 'SPECTRUM2'
Selects channel'SPECTRUM2'.
MMEM:STOR:TYPE CHAN
Specifies that channel data is to be stored.