Remote Commands in the Receiver Application
R&S
®
ESW
493User Manual 1177.6298.02 ─ 08
TTL
GPIB connection with TTL synchronization (if available; for most
Rohde&Schwarz devices)
*RST: GPIB
Example:
SYST:COMM:RDEV:GEN:LINK TTL
Selects GPIB + TTL interface for generator operation.
Manual operation: See "TTL Handshake" on page 160
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE <Type>
This command selects the type of external generator.
For a list of the available generator types see "Overview of Supported Generators"
on page 154.
Suffix:
<gen>
.
Parameters:
<Name> <Generator name as string value>
*RST: SMU02
Example: //Select an external generator
SYST:COMM:RDEV:GEN:TYPE 'SMW06'
Manual operation: See "Generator Type" on page 160
SYSTem:COMMunicate:TCPip:RDEVice:GENerator<gen>:ADDRess <Address>
Configures the TCP/IP address for the external generator.
Suffix:
<gen>
.
Parameters:
<Address> TCP/IP address between 0.0.0.0 and 0.255.255.255
*RST: 0.0.0.0
Example:
SYST:COMM:TCP:RDEV:GEN:ADDR 130.094.122.195
Manual operation: See "GPIB Address/TCPIP Address / Computer Name"
on page 160
15.6.1.4 User Port Configuration (AUX Port)
INPut<ip>:UPORt:STATe.................................................................................................494
INPut<ip>:UPORt[:VALue]...............................................................................................494
OUTPut<up>:UPORt[:VALue]..........................................................................................494
OUTPut<up>:UPORt:STATe............................................................................................495