EasyManua.ls Logo

Rohde & Schwarz ESW Series - Page 494

Rohde & Schwarz ESW Series
759 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands in the Receiver Application
R&S
®
ESW
494User Manual 1177.6298.02 ─ 08
INPut<ip>:UPORt:STATe <State>
This command toggles the control lines of the user ports for the AUX PORT connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S ESW.
See the R&S ESW Getting Started manual for details.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State> ON | 1
User port is switched to INPut
OFF | 0
User port is switched to OUTPut
*RST: 1
Manual operation: See "User port configuration" on page 145
INPut<ip>:UPORt[:VALue]
This command queries the control lines of the user ports.
For details see OUTPut<up>:UPORt[:VALue] on page 494.
Suffix:
<ip>
.
1 | 2
irrelevant
Return values:
<Level> bit values in hexadecimal format
TTL type voltage levels (max. 5V)
Range: #B00000000 to #B00111111
Example:
INP:UPOR?
//Result: #B00100100
Pins 5 and 7 are active.
Manual operation: See "User port configuration" on page 145
OUTPut<up>:UPORt[:VALue] <Value>
This command sets the control lines of the user ports.
The assignment of the pin numbers to the bits is as follows:
Bit 7 6 5 4 3 2 1 0
Pin N/A N/A 5 3 4 7 6 2
Bits 7 and 6 are not assigned to pins and must always be 0.
The user port is written to with the given binary pattern.
Configuration

Table of Contents