Remote Commands in the Receiver Application
R&S
®
ESW
486User Manual 1177.6298.02 ─ 08
Manual operation: See "Input Selection" on page 122
15.6.1.2 Preselector Configuration
INPut<ip>:ATTenuation:LIMiter[:STATe]............................................................................ 486
INPut<ip>:PRESelection:FILTer:NOTCh<notch>[:STATe]....................................................486
INPut<ip>:PRESelection:FILTer:SPLit[:STATe]...................................................................487
INPut<ip>:PRESelection[:STATe].....................................................................................487
INPut<ip>:ATTenuation:LIMiter[:STATe] <State>
This command turns the pulse limiter on and off.
The pulse limiter is an additional protection mechanism for the second RF input that
attenuates high level pulses.
Suffix:
<ip>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON
Example: //Turn on pulse limiter
INP:ATT:LIM ON
Manual operation: See "Pulse Limiter" on page 149
INPut<ip>:PRESelection:FILTer:NOTCh<notch>[:STATe] <State>
This command turns a preselecting notch filter on and off.
Suffix:
<ip>
.
irrelevant
<notch> 1..2
Selects the filter.
• <1> Selects the notch filter suppressing signals from 2.4 GHz
to 2.5 GHz.
• <2> Selects the notch filter suppressing signals from
2.725 GHz to 2.875 GHz.
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Turn on notch filter for frequencies from 2.725 GHz to
2.875 GHz.
INP:PRES:FILT:NOTC2 ON
Manual operation: See "Preselector Filter Settings" on page 150
Configuration