EasyManua.ls Logo

Rohde & Schwarz ESW Series - Page 540

Rohde & Schwarz ESW Series
759 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands in the Receiver Application
R&S
®
ESW
540User Manual 1177.6298.02 ─ 08
CALCulate<n>:MARKer<m>[:STATe] <State>
This command turns markers on and off. If the corresponding marker number is cur-
rently active as a delta marker, it is turned into a normal marker.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK3 ON
Switches on marker 3.
Manual operation: See "Marker State" on page 231
See "Marker Type" on page 232
See "Select Marker" on page 232
CALCulate<n>:MARKer<m>:TRACe <Trace>
This command selects the trace the marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Trace>
Example: //Assign marker to trace 1
CALC:MARK3:TRAC 2
Manual operation: See "Assigning the Marker to a Trace" on page 232
CALCulate<n>:MARKer<m>:X <Position>
This command moves a marker to a specific coordinate on the x-axis.
If necessary, the command activates the marker.
If the marker has been used as a delta marker, the command turns it into a normal
marker.
Suffix:
<n>
.
Window
Analysis

Table of Contents

Related product manuals