Remote Commands in the Receiver Application
R&S
®
ESW
631User Manual 1177.6298.02 ─ 08
For details on the file format see Chapter 13.3.3, "Reference: Transducer Factor File
Format", on page 327.
Suffix:
<n>
.
irrelevant
Parameters:
<FileName> Name of the transducer factor to be exported.
<TransdName> Name of the transducer factor to be exported.
Example:
MMEM:STOR:TFAC 'C:\TEST', 'Transducer1'
Stores the transducer factor named "Transducer1" in the file
TEST.CSV.
[SENSe:]CORRection:SWITch:ADDRess
<NetworkAddress>
This command defines the network address of the switch matrix.
Parameters:
<NetworkAddress> String containing the network address of the switch matrix.
Make sure to add the type of network protocol to the string.
Example:
CORR:SWIT:ADDR 'TCPIP::192.0.2.0::INSTR'
CORR:SWIT:ADDR 'TCPIP::192.0.2.0::HISLIP'
CORR:SWIT:ADDR 'GPIB::20::INSTR'
Connects to a device with the corresponding network addresses.
Manual operation: See "Connection between R&S ESW to the RF switch"
on page 325
[SENSe:]CORRection:SWITch:COMMent <Comment>
This command defines a comment for a dataset that controls a switch matrix.
Parameters:
<Comment> String containing the comment.
Example:
CORR:SWIT:COMM 'This is a comment.'
Defines a comment.
Manual operation: See "Dataset name and file name" on page 325
[SENSe:]CORRection:SWITch:DEFault[:COMMand] <Command>
This command defines a remote command that is sent to the switch matrix before all
other commands.
This is useful, for example, to configure the switch matrix into a predefined state.
Parameters:
<Command> String containing the command.