Remote Commands in the Receiver Application
R&S
®
ESW
640User Manual 1177.6298.02 ─ 08
<FileName> String containing the name of the transducer factor.
Note that you can assign up to eight transducer factors to a par-
ticular frequency range. In that case, add additional strings con-
taining the file name after the first one separated by comma.
Example: See Chapter 15.9.4, "Transducers", on page 629 .
[SENSe:]CORRection:TSET:SELect <FileName>
This command creates or selects a transducer set.
Parameters:
<FileName> String containing the name of the transducer set.
If the name does not exist yet, the R&S ESW creates a trans-
ducer set by that name.
Example: See Chapter 15.9.4, "Transducers", on page 629 .
[SENSe:]CORRection:TSET:UNIT <Unit>
This command selects the unit of a transducer set.
Note that the unit of all transducer factors in a transducer set must be the same or in
relative terms (dB).
Before you can use the command, you have to select a transducer set with [SENSe:
]CORRection:TSET:SELect.
Parameters:
<Unit> String containing one of the following units:
DB | DBM | DBUV | DBUV_M | DBUA | DBUA_M | DBPW |
DBPT
*RST: DB
Example: See Chapter 15.9.4, "Transducers", on page 629 .
[SENSe:]CORRection:TSET:INPut<rf>[:STATe] <State>
[SENSe:]CORRection:TSET[:STATe] <State>
This command turns a transducer set on and off.
Before you can use the command, you have to select a transducer set with [SENSe:
]CORRection:TSET:SELect.
Parameters:
<State> ON | OFF
*RST: OFF
Example: See Chapter 15.9.4, "Transducers", on page 629 .
General Instrument Setup