Remote Commands in the Receiver Application
R&S
®
ESW
639User Manual 1177.6298.02 ─ 08
<SizeFileN>
Size of the file in bytes.
Example:
CORR:TSET:CAT?
would return, e.g.
3298,34482896896,'SET1,1520','SET2,1756'
Usage: Query only
[SENSe:]CORRection:TSET:COMMent <Comment>
This command defines a comment for a transducer set.
Before you can use the command, you have to select a transducer set with [SENSe:
]CORRection:TSET:SELect.
Parameters:
<Comment> String containing the comment.
Example: See Chapter 15.9.4, "Transducers", on page 629 .
[SENSe:]CORRection:TSET:DELete
This command deletes a transducer set.
Before you can use the command, you have to select a transducer set with [SENSe:
]CORRection:TSET:SELect.
Example:
CORR:TSET:SEL 'Transducer Set'
CORR:TSET:DEL
Deletes the transducer set.
Usage: Event
[SENSe:]CORRection:TSET:RANGe<range> <Frequency>, <Frequency>,
<FileName>
This command selects a set of transducer factors used for a particular frequency
range.
Ranges 1 to 10 must be sent in ascending order.
Before you can use the command, you have to select a transducer set with [SENSe:
]CORRection:TSET:SELect.
Suffix:
<range>
.
1...10
Selects the range.
Parameters:
<Frequency> Defines the start frequency of the frequency range.
<Frequency> Defines the stop frequency of the frequency range.
General Instrument Setup