Remote Commands in the Receiver Application
R&S
®
ESW
671User Manual 1177.6298.02 ─ 08
For details on error queues see Chapter 14.1.7, "Status Reporting System",
on page 377.
Usage: Query only
SYSTem:FIRMware:UPDate <Directory>
This command starts a firmware update using the *.msi files in the selected directory.
The default path is D:\FW_UPDATE. The path is changed via the MMEMory:COMMent
command. To store the update files the MMEMory:DATA command is used.
Only user accounts with administrator rights can perform a firmware update.
Setting parameters:
<Directory>
Example:
SYST:FIRM:UPD 'D:\FW_UPDATE'
Starts the firmware update from directory "D:\FW_UPDATE".
SYSTem:FORMat:IDENt
<IDNFormat>
This command selects the response format to the *IDN? query.
Parameters:
<IDNFormat> LEGacy
Format is compatible to R&S FSP/FSU/FSQ/FSG family.
NEW | FSL
R&S ESW format
Format is also compatible to the R&S FSL and R&S FSV family
*RST: not reset!
Example:
SYST:FORM:IDEN LEG
Adapts the return value of *IDN? to the R&S FSP/FSU/FSQ fam-
ily.
Manual operation: See "*IDN Format" on page 402
SYSTem:PRESet:COMPatible <OpMode>
This command defines the operating mode that is activated when you switch on the
R&S ESW or press the [PRESET] key.
Parameters:
<OpMode> SANalyzer
Defines Signal and Spectrum Analyzer operating mode as the
presetting.
RECeiver
Selects the receiver application as the default application
(default value).
Manual operation: See "Preset Mode" on page 339
General Instrument Setup