Remote Commands in the Receiver Application
R&S
®
ESW
666User Manual 1177.6298.02 ─ 08
*RST: OFF
Manual operation: See "Language" on page 403
See "Wideband" on page 405
SYSTem:REVision:FACTory
Resets the response to the REV? query to the factory default value.
For example, after a user string was defined using the SYSTem:REVision[:STRing]
on page 666 command. (REV? query available for HP emulation only, see SYSTem:
LANGuage on page 664.)
Example:
Define the system language:
SYST:LANG '8563E'
Set the response back to factory setting:
SYS:REV:FACT
Query the revision:
REV?
Response:
920528
Usage: Event
Manual operation: See "Resetting the Factory Revision" on page 405
SYSTem:REVision[:STRing] <Name>
Sets the response to the REV? query to the defined string (HP emulation only, see
SYSTem:LANGuage on page 664).
Parameters:
<Name>
Example: Define the system language:
SYST:LANG '8563E'
Query the revision:
REV?
Response:
920528
Set the response to 'NewRevision':
SYST:REV:STR 'NewRevision'
Query the response:
SYST:REV:STR?
Response:
NewRevision
Manual operation:
See "Revision String" on page 405