Remote Commands in the Receiver Application
R&S
®
ESW
664User Manual 1177.6298.02 ─ 08
Parameters:
<String> String containing the description of the instrument.
Manual operation: See "Identification String" on page 401
SYSTem:KLOCk <State>
This command activates the local lockout (remote control) or returns to the local mode.
Parameters:
<State> ON
LLO (local lockout)
OFF
GTL (go to local)
*RST: OFF
Example:
SYST:KLOK ON
Activates LLO (remote control)
Manual operation: See "Local" on page 408
SYSTem:LANGuage <Language>
This command selects the system language.
For details see Chapter 14.2, "GPIB Languages", on page 395.
Parameters:
<Language> "SCPI" | "8560E" | "8561E" | "8562E" | "8563E" | "8564E" |
"8565E" | "8566A" | "8566B" | "8568A" | "8568A_DC" | "8568B" |
"8568B_DC" | "8591E" | "8594E" | "71100C" | "71200C" |
"71209A" | "PSA89600" | "PSA" | "PXA" | "FSP" | "FSU" |
"FSQ" | "FSV" | "FSEA" | "FSEB" | "FSEM" | "FSEK"
*RST: SCPI
Example:
SYST:LANG 'PSA'
Emulates the PSA.
Manual operation: See "Language" on page 403
Note: If you use "PSA89600", you must switch to an HP language first before return-
ing to SCPI (in remote operation only). For the identical language "PSA", this inter-
mediate step is not necessary.
SYSTem:LXI:INFO
This command queries the LXI settings.
Return values:
<LXIInfo> <current version> | <LXI class> | <Computername> |
<MAC adress> | <IP adress> | <Auto MDIX>
String containing the current LXI parameters.
• <version>
General Instrument Setup