Remote Commands in the Receiver Application
R&S
®
ESW
669User Manual 1177.6298.02 ─ 08
DIAGnostic:SERVice:VERSinfo?
This command queries information about the hardware and software components.
Return values:
<Information> String containing the version of hardware and software compo-
nents including the types of licenses for installed options.
Example:
DIAG:SERV:VERS?
Queries the version information.
Response:
Instrument Firmware |1.00,
BIOS |R&S ANALYZER BIOS V1.70-3-14-2 IPC11,
Image Version |1.6.0,
Device Installation Version |1.3.0,
PCIE-FPGA |8.03,
SA-FPGA |10.05,
MB-FPGA |2.1.3.0,
SYNTH-FPGA |3.12.1.0,
REF-FPGA |3.4.0.0,
MWC-FPGA |3.4.0.0,
Data Sheet Version |06.00,
Time Control Management ||active,
RF Preamplifier B24||,
Real-Time Analysis K55||permanent
Usage:
Query only
SYSTem:ERRor:CLEar:ALL
This command deletes all contents of the "System Messages" table.
Example:
SYST:ERR:CLE:ALL
SYSTem:ERRor:EXTended? <MessageType>[, <ChannelName>]
This command queries all system messages, or all messages of a defined type, dis-
played in the status bar for a specific channel (application).
Note: This command queries the strings displayed for manual operation. For remote
programs, do not define processing steps depending on these results. Instead, query
the results of the STATus:QUEStionable:EXTended:INFO status register, which
indicates whether messages of a certain type have occurred (see "STATus:QUEStiona-
ble:EXTended:INFO Register" on page 386).
Parameters:
<MessageType> ALL | INFO | WARNing | FATal | ERRor | MESSage
<ChannelName> String containing the name of the channel.
The parameter is optional. If you omit it, the command works for
the currently active channel.