Remote Commands in the Receiver Application
R&S
®
ESW
442User Manual 1177.6298.02 ─ 08
INITiate<mt>:EMITest
This command initiates an automated test sequence.
The sequence consists of a scan, a peak search and a final measurement.
When you are running a continuous measurement, you have to stop the scan deliber-
ately (with ABORt) before the peak search is performed. For a single measurement,
the R&S ESW automatically performs the peak search after number of scans defined
by the scan count has been performed.
Suffix:
<mt>
.
irrelevant
Example: //Start the test sequence
INIT2:EMIT
Usage: Event
Manual operation: See "Performing single measurements" on page 96
INITiate<mt>:FMEasurement
This command initiates a final measurement based on the peak list.
Suffix:
<mt>
.
irrelevant
Example: //Start the final measurement
INIT2:FME
Usage: Event
Manual operation: See "Performing single measurements" on page 96
INITiate<mt>[:IMMediate]
The command initiates a new measurement.
For a single measurement, the R&S ESW stops measuring when it has reached the
end frequency. When you start a continuous measurement, it stops only if you abort it
deliberately.
If you are using trace modes MAXHold, MINHold and AVERage, previous results are
reset when you restart the measurement.
Single measurements
Synchronization to the end of the measurement is possible with *OPC, *OPC? or
*WAI.
Continuous measurements
Synchronization to the end of the measurement is not possible.
It is thus recommended to use a single measurement for remote controlled mea-
surements, because results like trace data or markers are only valid after synchro-
nization.