Remote Commands in the Receiver Application
R&S
®
ESW
451User Manual 1177.6298.02 ─ 08
Example: //Query bargraph peak level
DISP:BARG:LEV:UPP?
would return, e.g.
110
Usage: Query only
DISPlay:BARGraph:PHOLd:RESet
This command resets the bargraph max hold.
Example: //Reset the max hold value
DISP:BARG:PHOL:RES
Usage: Event
Manual operation: See "Bargraph Max Hold" on page 100
DISPlay:BARGraph:PHOLd[:STATe]
<State>
This command turns the bargraph max hold function on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Turn on the bargraph max hold
DISP:BARG:PHOL ON
Manual operation: See "Bargraph Max Hold" on page 100
DISPlay:BARGraph:TCOupling[:STATe] <State>
This command couples or decouples the bargraph detector and trace detector used for
the scan.
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Couple the type and color of bargraph and scan trace
DISP:BARG:TCO ON
Manual operation: See "Couple to Scan Trace" on page 100
[SENSe:]DETector<t>:RECeiver[:FUNCtion] <Detector>...
This command selects the detector for the bargraph measurement.