Remote Commands in the Receiver Application
R&S
®
ESW
508User Manual 1177.6298.02 ─ 08
Example: //Select low noise auto ranging mode
INP:ATT:AMOD LOWN
Manual operation: See "Auto Range" on page 120
INPut<ip>:ATTenuation:AUTO <State>
This command turns automatic determination of the attenuation level on and off.
When you turn it on, the R&S ESW selects an attenuation that results in a good signal-
to-noise ratio without overloading the RF input.
Suffix:
<ip>
.
irrelevant
Parameters:
<State> ON | OFF
ON
Receiver: Turns on the auto ranging feature.
OFF
Receiver: Turns off the auto ranging feature.
*RST: ON
Example: //Turn on auto ranging
INP:ATT:AUTO ON
Manual operation: See "Auto Range" on page 120
See "Attenuation" on page 121
INPut<ip>:ATTenuation:PROTection[:STATe] <State>
This command turns the availability of attenuation levels of 10 dB or less on and off.
Suffix:
<ip>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: 1
Example: //Turn on input protection
INP:ATT:PROT ON
Manual operation: See "10 dB Minimum Attenuation" on page 173
INPut<ip>:GAIN:AUTO <State>
This command includes and excludes the preamplifier from the auto ranging feature.