Remote Commands in the Receiver Application
R&S
®
ESW
527User Manual 1177.6298.02 ─ 08
Parameters:
<x1> Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
<y1> Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
<x2> Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
<y2> Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
Manual operation: See "Single Zoom" on page 196
DISPlay[:WINDow<n>][:SUBWindow<w>]:ZOOM[:STATe] <State>
This command turns the zoom on and off.
Suffix:
<n>
.
Window
<w> subwindow
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DISP:ZOOM ON
Activates the zoom mode.
Manual operation: See "Single Zoom" on page 196
See "Restore Original Display" on page 196
See " Deactivating Zoom (Selection Mode)" on page 196
Analysis