Example: Sequencer procedure
Assume the following active channel definition:
Tab name Application Sweep mode Sweep count
Spectrum Spectrum Cont. Sweep 5
Spectrum 2 Spectrum Single Sweep 6
Spectrum 3 Spectrum Cont. Sweep 2
IQ Analyzer IQ Analyzer Single Sweep 7
For Single Sequence, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer
For Continuous Sequence, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
...
For Channel-defined Sequence, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 2 x Spectrum 3,
5x Spectrum, 2 x Spectrum 3,
...
RUN SINGLE/RUN CONT and Single Sweep/Sweep Continuous keys
While the Sequencer is active, the [RUN SINGLE] and [RUN CONT] keys control the
Sequencer, not individual sweeps. [RUN SINGLE] starts the Sequencer in single
mode, while [RUN CONT] starts the Sequencer in continuous mode.
Running a Sequence of Measurements