Performing Configuration Tasks
R&S
®
SGS100A
160User Manual 1173.9105.02 ─ 12
Tip: Calibration error. If the "Info" line shows the message "Calibration Error",
select "SGMA-GUI main panel > Instrument Name > Setup > Internal Adjustments
> Adjust All" to trigger internal adjustment.
7. If necessary, install the new R&S SGMA-GUI.
For detailed description, refer to Chapter 2.3.1, "Installing the R&S SGMA-GUI
Software", on page 23
8. If the instrument and the controller/PC are connected over the PCIe interface and
the external PC does not support hot-plugging, restart the external PC.
Firmware update through a session control protocol (SCP)
If an update to a new firmware version is required for both the R&S SGMA-GUI and the
instrument, it is mandatory that the new instrument's firmware is installed before the
update of the R&S SGMA-GUI to the new version.
1. Connect the R&S SGS and a Windows PC to the same network.
2. On the PC, open a windows explorer window.
3. To connect to the R&S SGS, enter the name of the instrument or its IP address in
the windows taskbar.
4. Enter the user name and password to connect to the R&S SGS. The default user
name is instrument and the password is instrument.
A folder opens, containing the share and the update folder.
5. Open the update folder and copy the new firmware update file in it.
The update starts automatically.
Unsuccessful or erroneous firmware update
An erroneous or unsuccessful installation of firmware update package is indicated by a
combination of one orange and red LEDs on the front panel.
Refer to the service manual for a description of the displayed error code or contact the
customer support center.
How to Install a New Firmware Version on the Instrument