5. Application Operation
5-94
CBM-20A/20Alite
Method Parameters for the Column Oven
CTO-30AS
CTO-30A
CTO-20A, CTO-20AC
CTO-10Avp, CTO-10ACvp, CTO-10ASvp, CTO-10A, CTO-10AC
Parameter Setting range Minimum unit Default setting Remark
Oven Temperature (C) 0, 4 to 85 1 40 0 means OFF.
Maximum Temperature (C) 5 to 90 1 90
Parameter Setting range Minimum unit Default setting Remark
Oven Temperature (C) 0, 4 to 150 1 40 0 means OFF.
Maximum Temperature (C) 5 to 160 1 160
Heat Compensation
AUTO
MANUAL
AUTO
AUTO: The flow rate of pump A set
in the "Method Parameter" window is
set as the heat compensation flow.
MANUAL: The value set for [Heat
Compensation Flow] is set.
Heat Compensation Flow
(mL/min)
0 to 10.0000 0.0001 0
Setting is available when "MANUAL"
is selected for [Heat Compensation].
Valve Position
2 Position Valve R
2 Position Valve L
0, 1 0
Set the position of the 2-position
valves R and L.
Valve Position
6 Position Valve R
6 Position Valve L
1, 2, 3,
4, 5, 6
1
Set the position of the 6-position
valves R and L.
Parameter Setting range Minimum unit Default setting Remark
Oven Temperature (ºC) 0, 4 to 85 1 40 0 means OFF.
Maximum Temperature (ºC) 5 to 90 1 90
Valve Position
2 Position Valve R
2 Position Valve L
0, 1 0
Sets the position of 2-position valves
R and L.
Valve Position
6 Position Valve R
6 Position Valve L
1, 2, 3,
4, 5, 6
1
Sets the position of 6-position valves
R and L.
Parameter Setting range Minimum unit Default setting Remark
Oven Temperature (ºC) 0, 4 to 80 1 40 0 means OFF.
Maximum Temperature (ºC) 5 to 85 1 85