5.13 List of Time-program Commands
5-133
CBM-20A/20Alite
5
RF-20A/20Axs
1. Cannot be used with the RF-20A.
Module Function Description
Setting
range
Minimum
unit
Default
setting
Remark
DetectorA
DetectorB
EX.WAV
Excitation wavelength
(nm)
0
200 to 900
1350
Can be set to [0]
only if [Single] is
selected as the
wavelength mode.
EM.WAV
Emission wavelength
(nm)
0
200 to 900
1450
Can be set to [0]
only if [Single] is
selected as the
wavelength mode.
EX.WAV2
Excitation
wavelength2 (nm)
200 to 900 1 350
Settings for dual
mode
EM.WAV2
Emission wavelength2
(nm)
200 to 900 1 450
Settings for dual
mode
RANGE2 Recorder range2 0 to 9 1 1
Settings for single
mode
RESP Response 0 to 10 1 4
GAIN Gain
1:x1
2:x4
3:x16
12
SENSE Sensitivity
1:HIGH
2:MID
3:LOW
12
LAMP Lamp selection
0:OFF
1:ON
11
CELL.T
(See note 1.)
Cell temperature (ºC) 0, 4 to 40 1 30 0 to OFF
MARK
Marker output to
recorder terminal
ZERO
Execution of auto-
zero