5. Application Operation
5-134
CBM-20A/20Alite
RF-10AXL/10A
Module Function Description
Setting
range
Minimum
unit
Default
setting
Remark
DetectorA
DetectorB
EX.WAV
Excitation wavelength
(nm)
0
200 to 900
1350
EM.WAV
Emission wavelength
(nm)
0
200 to 900
1450
RANGE Recorder range 0 to 9 1 1
RESP Response
1:0.1
2:0.5
3:1.5
4:3.0
13
GAIN Gain
1:x1
2:x4
3:x16
12
SENSE Sensitivity
1:HIGH
2:MID
3:LOW
12
With the RF-10A,
available settings
are 1:HIGH and
2:LOW.
LAMP Lamp selection
0:OFF
1:ON
11
DetectorA
DetectorB
MARK
Marker output to
recorder terminal
ZERO
Execution of auto-
zero