6.2 Error Message
6-7
CBM-20A/20Alite
6
• Settings made in the "Configuration Parameter" window and the "Configuration" tab
page (including settings for the system controller's serial interface)
Component and user information includes the following data.
• Component information such as system and group names
• User names, user IDs, passwords, and access levels
• Results and execution dates of system checks
Data that has been lost cannot be restored.
• It is recommended that method and sequence files are saved by executing [Save All
Parameters] in the "Utility" window.
• The creation of backup files for component and user information is not possible.
Ensure that users make records of settings as appropriate.
• If there is a CBM operating in the same group, user information can be copied from
that CBM.
6.2.2 Operating Errors
If an error occurs in the system controller or in a connected component while the user is
performing window operations, an error message is displayed. If LC-30A/20A-series
components are connected, the buzzer sounds and the error is displayed at one of these
components.
In the case of the Nexera UC series instruments (LC-30ADSF, SFE-30A, and SFC-
30A), operations using Internet Explorer are not guaranteed and no error message will
be displayed.
Error Window