6. Troubleshooting
6-8
CBM-20A/20Alite
Error
The 3 types of error are described in the following table.
Warning
No processing is performed by the system controller. It becomes possible, however, to make
settings from the operating panel of the displayed component. Control is not possible from the
system controller.
Display item Description
Error code A 6-digit alphanumeric error code is displayed.
Component name
The name of the component for which the error occurred is
displayed.
Error message A description of the error is displayed.
All-clear button Click to clear all errors except fatal errors.
Clear button
Click to clear the currently displayed (non-fatal) error. If there is
more than one error, the next error is displayed.
Logout button Click to log out.
Error Message Description
Warning Urges the user to exercise caution.
Error Indicates an error in the system.
Fatal error
Indicates a problem related to hardware. The message [Turn OFF the
power.] is displayed.
Instrument Error Code Error Message
Pump
0x0033 Operation mode is local. Set the operation mode to remote.
0x81D0 Mobile phase A is low.
0x81D1 Mobile phase B is low.
0x81D2 Mobile phase C is low.
0x81D3 Mobile phase D is low.
0x81E7 Cannot write parameters in EEPROM.
0x81F1
A seal malfunction has occurred in the online degassing unit
vacuum line.
Autosampler
0x0094 Operation mode is local. Set the operation mode to remote.
0x8541 Rack on rack plate L is not adjusted.
0x8542 Rack on rack plate M is not adjusted.
0x8543 Rack on rack plate R is not adjusted.
0x8544 Invalid rack ID.
0x8545 Racks with same rack ID are set.