6.2 Error Message
6-9
CBM-20A/20Alite
6
Autosampler
0x8546
The temperature control of rack plate L has been turned off
because no rack has been set for more than 30 minutes. If the
rack is set, the temperature control will be on again.
0x8547
The temperature control of rack plate M has been turned off
because no rack has been set for more than 30 minutes. If the
rack is set, the temperature control will be on again.
0x8548
The temperature control of rack plate R has been turned off
because no rack has been set for more than 30 minutes. If the
rack is set, the temperature control will be on again.
0x85EB
It is necessary to set all racks on the stacks at the rack-
changer. Check all racks are set.
0x85EC The stack is open. Close the stack in the rack-changer soon.
0x85ED
The temperature control has been off because stack has been
opened for more than 30 minutes. If the stack is closed, the
temperature control will be on again.
0x85EE
As the stack has been closed, the temperature control has
been on again.
0x85EF
No rack is detected. Check the rack on the stack. If the rack is
not set and analysis is executed, the sequence will be
stopped.
0x85F0 The reserved warning code 1 has been detected.
0x85F1
Full rack is detected. Remove the duplex rack on the stack or
set the dummy rack. If the rack is not changed, the sequence
will be stopped.
0x8B00 Operation mode is local. Set the operation mode to remote.
0x8B02 Autosampler and SFE Unit are connected at the same time.
0x8B45 Low set temperature.
Column oven
0x0075 Operation mode is local. Set the operation mode to remote.
0x0077 Low set temperature.
0x0078 High set temperature.
0x8632 Cannot write in CMD.
Detector
0x0046 Operation mode is local. Set the operation mode to remote.
0x0049
Auto zero level is out of range. Confirm the installation of the
cell, no bubble in the cell, and the settings of SENS and GAIN
parameters.
0x0050 No scan data
0x87D0 Low set temperature.
0x8721 Low set temperature.
PDA Detector
0x8952 LAN timeout error has occurred. (PC)
0x8954 LAN timeout error has occurred. (CBM)
0x89D2 Temperature is increasing inside the instrument.
0x89FB Undefined warning.
Instrument Error Code Error Message