8012428/YWL2/3-0/2016-08| SICKO P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER T
Subject to change without notice
5 MAINTENANCE
Fig. 65: Cleaning the optical interfaces on the sender/receiver unit
Lay the toothed belt on the drive axis, press the tension spring together and push
pivoted shutter back onto the axis.
To start the function check, connect to file “MCU”, select subdirectory “Adjustment /
Manual function check” and actuate “Start Manual Function Check”.
Fig. 66: SOPAS ET menu: MCU/Adjustment/Function Check - Manual
In SOPAS ET, connect to device file “DH T100” or “DH T200”, select directory “Diagnosis
/ Check values” and check the contamination value.
Zero point reflector,
rear side
Pivoted shutter cover
Glass pane (both sides)
Zero point reflector
Axis
Pivoted shutter
Tension spring
Sender optics
The function check can also be triggered via the LC-Display on the MCU (see “Menu
structure”, page 87).