General Operation
7.11 Open Channel Monitoring (OCM)
SITRANS LUT400
126 Operating Instructions, 07/2019, A5E33329501-AC
7.11.4.3 Parshall Flume
Note
C = Converging Dimension.
Zero head
Converging dimension C
2/3 C
Flow
Transducer*
* The transducer must be above the maximum head by at least the blanking value (see Near
Range (2.2.4.) (Page 172)).
Application Information
Sized by throat width
Set on solid foundation
For rated flows under free flow conditions, the head is measured at 2/3 the length of the
converging section from the beginning of the throat section.
Parameter
Value
Primary Measuring Device (PMD) (2.15.1.) (Page 230)
Exponential Devices
Flow Exponent (2.15.3.2.) (Page 231)
1.5221.607
a
Maximum Head (2.15.3.3.) (Page 232)
Maximum Flow at 20 mA (2.15.3.4.) (Page 232)
K Factor (2.15.4.1.) (Page 235)
b
a
Typical Flow Exponent range for Parshall Flume; consult your flume documentation.