Commissioning
6.5 Requesting an Echo Profile
SITRANS LUT400
78 Operating Instructions, 07/2019, A5E33329501-AC
6.5 Requesting an Echo Profile
1
Agorithm
2
BLF (Best of First or Largest Echo)
Distance from transducer face to target
1)
See Confidence (3.2.9.2.) (Page 245)
2)
See Algorithm (2.12.2.1.) (Page 220)
● In PROGRAM mode, navigate to: Main Menu > Diagnostics (3.2.) > Echo Profile (3.2.1.).
● Press RIGHT arrow
to request a profile.
● Use UP
or DOWN arrow to scroll to an icon. When an icon is highlighted, that
feature becomes active.
● To move a cross-hair, press RIGHT
arrow to increase the value, LEFT arrow to
decrease.
● To Zoom into an area, position the intersection of the cross-hairs at the center of that
area, select Zoom, and press RIGHT arrow . Press LEFT arrow to Zoom out.
● To update the profile, select Measure and press RIGHT
arrow .
● To return to the previous menu, select Exit then press RIGHT
arrow .
6.6 Device Address
Setting a device address is not necessary for local operation, but must be set if configuring
the SITRANS LUT400 for use on a HART network. See Device Address (4.1.) (Page 260).