SITRANS LUT400
284 Operating Instructions, 07/2019, A5E33329501-AC
Technical data
12
12.1 Power
AC model
100-230 V AC ±15%, 50 / 60 Hz, 36 VA (10W)
1)
Fuse: 5 x 20 mm, Slow Blow, 0.25A, 250V
DC model
10-32 V DC, 10W
1)
Fuse: 5 x 20 mm, Slow Blow, 1.6A, 125V
1)
Power consumption is listed at maximum.
12.2 Performance
Range
0.3 to 60 m (1 to 196 ft), dependent on transducer
Accuracy (measured under Reference Conditions similar to IEC 60770-1)
Standard operation: ±1 mm (0.04") plus 0.17 % of distance
High accuracy OCM
1)
: ±1 mm (0.04"), within 3 m (9.84 ft) range
Resolution (measured under Reference Conditions similar to IEC 60770-1)
Standard operation: 0.1 % of range or 2 mm (0.08”), whichever is greater
High accuracy OCM
1)
: 1 mm (0.04")
2)
, within 3 m (9.84 ft) range
Reference operating conditions according to IEC 60770-1
ambient temperature +15 to +25 °C (+59 to +77 °F)
humidity 45% to 75% relative humidity
ambient pressure 860 to 1060 mbar g (86 000 to 106 000 N/m
2
g)
Temperature compensation
Range: -40 to +150 °C (-40 to +300 °F)
Source
Integral transducer sensor
TS-3 temperature sensor
Average (integral transducer and TS-3)
Programmable fixed temperature
Temperature error
Fixed
0.17 % per °C deviation from programmed value
Memory
512 kB flash EPROM
1.5 MB flash for data logging
1)
Power consumption is listed at maximum.
2)
A high accuracy configuration consists of the LUT440 (OCM) model using XRS-5
transducer, TS-3 temperature sensor, and a Low Calibration Point of 3 m or less. Under
severe EMI/EMC environments per IEC 61326-1 the DC powered device may have an
additional error increase of up to 0.5 mm.