Installing and Mounting
4.2 Mounting instructions
SITRANS LUT400
Operating Instructions, 07/2019, A5E33329501-AC
27
Note
Recommended mounting: directly to horizontal or vertical pipe. If alternate mounting
surface is used, it MUST be able to support four times the weight of the device.
Recommended pipe dimensions: maximum: 3" pipe, minimum: 3/4" pipe.
Recommended fastener sizes:
U-Bolts:
Maximum : 3" pipe size with M8 or 3/8" thread.
Minimum: 3/4" pipe size with M6 or 1/4" thread.
Hex Nuts:
M6 or 1/4" to M8 or 3/8".
Washer:
Maximum: 16 mm or 13/16" outside diameter.