Commissioning
6.3 Quick Start Wizards
SITRANS LUT400
62 Operating Instructions, 07/2019, A5E33329501-AC
High Calibration Point
Distance from Sensor Reference Point to High Calibration Point: usually process full level.
Low Calibration Point
Distance from Sensor Reference Point to Low Calibration Point: usually process empty level.
Default: 60.000
Response Rate
Sets the reaction speed of the device to measurement changes in the target range.
Response Rate can only be set through the Quick Start Wizard, and any changes made
to
Fill Rate per Minute (2.3.1.) (Page 174), Empty Rate per Minute (2.3.2.) (Page 174
Damping Filter (2.3.3.) (Page 174) parameters following the completion of the wizard will
supersede the Response Rate setting.
Response Rate always displays in m/minute.
Default: SLOW (0.1 m/min)
Use a setting just faster than the maximum filling or emptying rate (whichever is greater).
Slower settings provide higher accuracy, faster settings allow for more rapid level
fluctuations.
End of QS Level Wizard
For QS to be successful, all changes must be applied.
BACK, CANCEL, FINISH
(Display returns to 1.1 Quick Start menu when Quick Start is successfully completed or
cancelled. If CANCEL is selected, no changes are written to the device.)
To transfer Quick Start values to the device and return to Program menu, press DOWN
arrow
(Finish). Then press LEFT arrow three times to return to Measurement mode.