LC700 – User’s Guide
3.6
COMMUNICATION PORTS
Quantity 3 (P1, P2, P3)
Types
P1 – EIA - 232C (Point-to-Point) Modbus RTU
P2 – EIA - 485 (Multi-drop) Modbus RTU
P3 – EIA - 485 (Multi-drop) Modbus RTU
Connectors
DB9 Female for P1
Terminal Block (3 Points)
Operation Independent
Label See Modules and Accessories
Baud Rate/Address
P1 1200, 2400, 4800, 9600, 19.2K, 38.4K,
57.6K bps
P2/P3 as Modbus RTU: 1200, 2400, 4800,
9600, 19.2K, 38.4K, 57.6K, 115200K bps
Protocol
Modbus RTU (Slave) or Dedicated SMAR
Remote I/O
CPU ID ( Modbus Address )
1 to 200 assigned by the user (1 is the default
address)
Maximum Number of LC700 CPU’s per
Network
31
INTERNAL POWER
Provided by the IMB bus 5VDC, @ 240mA
Total Maximum Dissipation 1.20W
Power Indicator "+5VDC" Green LED
FAILURE CIRCUIT
Output Type
Relay, Normally Close (NC). Relay is OPEN
for normal operation
Contact Voltage Limits 20 – 115 VAC/VDC
Maximum Contact Current for 115 VAC 1.0 A (resistive); 0.5 A (inductive)
Maximum Contact Current for 24 VDC 1.0A (resistive); 0.5A (inductive)
Maximum Initial Contact Resistance
50 m
Status Display Red LED "Fail" (Contact Closed)
Relay Power Indicator
Green LED - ON if the relay circuit is powered
(contact closed)
Leakage 500 mA @ 100VAC
Over Load Protection Should be provided externally
Internal RC Protection Circuit
62 in series with 0.01 μF
Operate Time 10 ms maximum
Release Time 10 ms maximum
Electrical service life 100.000 operations min @ 0.5A ,115 VAC
External Power Indicator "Relay Coil Power" Green LED
External Power 8.3 mA @ 24 VDC
External Voltage Range
23-30 VDC (excessive voltage can burn the
relay coil)
Resistance
2880 (Coil)
Inductance 2.7 H (Coil)
Optical Isolation 5000Vac before the Relay’s isolation.