7-11
Troubleshooting Operator Manual 920011-649
Alarm sounds, and display shows:
FAULT 23
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 23 =
POWER FAILURE
1. Power failure occurred or POWER (ON/OFF) toggle switch
pressed during a cycle.
2. When power is restored, refer to S
ECTION 4.15 for
procedure.
Alarm sounds, and display shows:
FAULT 24
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 24 =
AIR PURGE RTD OVERHEAT
1. Air purge pre-filter needs changing, call STERIS.*
2. Air purge HEPA filter needs changing, call STERIS.*
3. If filters were changed and situation reoccurs, call
STERIS.*
Alarm sounds, and display shows:
FAULT 25
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 25 =
BOOT A: LOW PRESSURE
1. An endoscope of a diameter smaller than template is
processed without proper insertion tube sleeve. Use
insertion tube sleeve in control handle boot "A" as per
procedure (refer to S
ECTION 4.3).
2. An endoscope without a light cord is processed without
control handle boot plug. Use control handle boot plug in
control handle boot "A" as per procedure (refer to
S
ECTION 4.3).
3. Only one endoscope is processed and has been put in
control handle boot "B" instead of "A" (refer to S
ECTION 4.3).
4. Control handle boot is not latched. Latch and try again.
5. Clogged debris screen. Clean debris screen.
Continued...
*Service charges may be incurred. Consult your warranty for details.
FAULT POSSIBLE CAUSE AND CORRECTION