7-13
Troubleshooting Operator Manual 920011-649
Alarm sounds, and display shows:
FAULT 27
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 27 =
BOOT A: PRESSURE SENSOR DEFECTIVE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 28
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 28 =
BOOT B: PRESSURE SENSOR DEFECTIVE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 29
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 29 =
0.2 µm FILTER OVER PRESSURE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
FAULT POSSIBLE CAUSE AND CORRECTION