0R1!0+323+331+351+0.0+0.4+3.0<cr><lf>
0R3!0+0.15+20+0.0+0.0+0+0.0+0.0+0.0<cr><lf>
0R!0+178+288+001+15.5+27.4+38.5+23.9+35.0+1002.1+0.00+0+0.0+23.8<cr><lf>
7.3.12 Continuous Measurement with CRC (aRC)
Command format:
aRCx!
The device can be configured so that all the parameters can be requested instantly with the
command
aRC but a three-character CRC is added to the response data strings before
<cr><lf>.
Example (device address 0):
0RC3!0+0.04+10+14.8+0.0+0+0.0INy
7.4 NMEA 0183 v3.0 Protocol
This section lists the data query commands and data message formats for the NMEA 0183
v3.0 query and automatic protocols.
A two-character checksum (CRC) field is transmitted in all data request sentences.
More In
formation
‣
Sensor Configuration and Data Message Formatting (page 117)
‣
CR
C-16 Computation (page 179)
7.4.1 Device Address (?)
This command queries the address of the device on the bus.
C
ommand f
ormat:
?<cr><lf>
?
Device address query command
<cr><lf>
Command terminator
WXT530 Series User Guide M211840EN-D
102