Table 36 Error Messaging/Text Messages
Text message
identifier (in
NMEA 0
183
v3.0 protocol
only)
Text Message Interpretation and action
01 Unable to measure
error
T
he requested parameters are not activated in the message and
check the parameter section fields.
02 Sync/address error The device address in the beginning of the command is invalid.
Ask the device address with the ?! (SDI-12) or ?<cr><lf> (ASCII
and NMEA) command and retype the command with the correct
address.
03 Unknown cmd error The command is not supported, use the correct command format.
See 7. R
etrieving Data Messages (page 77).
04 Profile reset Check
sum error in configuration settings during power-up.
Factory settings used instead.
05 Factory reset Checksum error in calibration settings during power-up. Factory
set
tings used instead.
06 Version reset New software version in use.
07 Start-up Software reset. Program starts from the beginning.
08 Use chksum xxx Given checksum not correct for the command. Use the proposed
checksum.
09 Measur
ement reset The ongoing measurement of all the sensors interrupted and
start
ed from the beginning.
10 Rain reset The precipitation sensor counter reset.
11 Inty reset Precipitation sensor intensity counter reset.
If you use a WXT530 transmitter on an RS-485 bus with other polled devices, you must
disable error mes
saging with the command: 0SU,S=N<crlf>.
More Information
‣
T
r
oubleshooting (page 151)
‣
Sensor Configuration
and Data Message Formatting (page 117)
10.1.2 Rain and Wind Sensor Heating Control
The supervisor message aSU sho
w
s you continuously monitored information about rain and
wind sensor heating (heating temperature Th and heating voltage Vh).