Problem Possible causes Action(s)
The transmitter keeps
sending the message "T
X
Sync/address error".
The polling address and the transmitter
address do not match. The transmitter is
on an RS-485 bus with other polled
devices and Error Messages are
enabled.
Set correct address either for the
transmitter or to the polling request.
Disable the Error Messages with the
command aSU,S=N <cr><lf>.
mA messages cannot be
found.
No mA output or
dered. You must specify the mA option when
you place the order.
Analog input messages are
missing.
Analog input mes
sages are not enabled. Enable analog input messages. See
8.1.7.5 Enabling or Disabling Analog
Output (page 145).
Precipitation messages
missing.
No pr
ecipitation measurement in
models WXT534 and WXT532.
10.1 Self-Diagnostics
10.1.1 Error Messaging/Text Messages
The transmitter sends a text message when certain type of errors occur. The function works
in all communica
tion modes except in the SDI-12 mode. You can disable error messaging by
using the supervisor message aSU, S=N.
Examples:
0R1!0TX,Unable to measure error<cr><lf>
(reques
t of wind data while all the wind parameters were disabled from the wind message)
1XU!0TX,Sync/address error<cr><lf>
(wrong device address. Ask the correct address with ? or ?! command.)
0XP!0TX,Unknown cmd error<cr><lf>
0xUabc!0TX,Use chksum CCb<cr><lf>
(wr
ong checksum applied to the 0xU command)
Chapter 10 – Troubleshooting
153
sales@streamlinemeasurement.co.uk