0R3<cr><lf>
0R3,Rc=0.00M,Ri=0.0M<cr><lf>
Example (
SDI-12, device address 0):
Changing the counter reset mode (resets the precipitation counters):
0XRU,Z=M!0<cr><lf>
More In
formation
‣
Setting Fields (page 128)
8.1.4 Supervisor Message
8.1.4.1 
Checking the Settings (aSU)
Use this c
ommand to check the current supervisor settings.
Command f
ormat in ASCII and NMEA 0183:
aSU<cr><lf>
Command f
ormat in SDI-12:
aXSU!
a
Device address
SU
Supervisor settings command in ASCII and NMEA 0183
XSU
Supervisor settings command in SDI-12
<cr><lf>
Command terminator in ASCII and NMEA 0183
!
Command terminator in SDI-12
The response in ASCII and NMEA 0183:
aSU,R=[R],I=[I],S=[S],H=[Y]<cr><lf>
The r
esponse in SDI-12:
WXT530 Series User Guide M211840EN-D
132