Appendix E. Factory
Configurations
The factory configurations are read-only settings which cannot be modified. For each
set
tings command, the following information is shown:
• Command to retrieve the settings (ends to ! character)
• Example response from the transmitter
• Table describing the message contents
E.1 General Unit Settings
0XF!0XF,f=11111111&11100010,o=AAC1DB1A,c=A263,
i=HEL___,n=A3430012,2=2528,3=3512 <cr><lf>
Table 49 General Unit Settings
Field
Charac
ter
Field Name Description
f Factory options Selection of parameters
o Order code Ordering identity as delivered (10 characters)
c Calibration date Y=2003, A, B,…=2005, 2006, 1..52 = week, 1 … 7, weekday
i Info Factory signature (10 characters)
n Device s/n A,B,...=2005,2006..., 1..52 = week, 1 … 7 = weekday , 1 … 9999 = serial
number
2 2.5 V reference 2500 mV (default)
3 3.5 V reference 3500 mV (default)
E.2 Wind Configuration Settings
0WF!0WF,g=A,l=N,t=A,
0=273.00,1=273.01,2=273.00,3=273.00,4=273.00,5=273.00,a=45.1,b=50.2,u=54.9,v=63
.1,x=65.1,y=65.1<cr><l f>
Appendix E – Factory
Configurations
183