<cr><lf>
Command terminator
Example of the response (the parameter set is configurable):
0R1,Dn=236D,Dm=283D,Dx=031D,Sn=0.0M,Sm=1.0M, Sx=2.2M<cr><lf>
a
Device address
R1
Wind message query command
Dn
Wind direction minimum (D = degrees)
Dm
Wind direction average (D = degrees)
Dx
Wind direction maximum (D = degrees)
Sn
Wind speed minimum (M = m/s)
Sm
Wind speed average (M = m/s)
Sx
Wind speed maximum (M = m/s)
<cr><lf>
Response terminator
7.2.5 Pressure, Temperature and Humidity Data Message (aR2)
This command requests a pressure, temperature, and humidity data message.
C
ommand f
ormat:
aR2<cr><lf>
a
Device address
R2
Pressure, temperature and humidity message query command
<cr><lf>
Command terminator
Example of the response (the parameter set is configurable):
0R2,Ta=23.6C,Ua=14.2P,Pa=1026.6H<cr><lf>
a
Device address
R2
Pressure, temperature and humidity message query command
Ta
Air temperature (C = °C)
WXT530 Series User Guide M211840EN-D
84