E.4 Rain Configuration Settings
0RF!0RF,p=1.0,n=3.0,d=N,f=0<cr><lf>
Table 52 Rain Configuration Settings
Field
Character
Field Name Description
p, n Positive and
negativ
e gain
0.1 … 25.5 (p=1.0, n=1.0)
d Bypass all hits Y=Enabled, N= Disabled (default)
f Wind filter bypass 0,1 … 4 (0=wind depended, 1,2,3,4=threshold level)
E.5 Supervisor Settings
0SF!0SF,t=19.8,b=17159,l=-4.0,m=0.0,h=4.0<cr><lf>
Table 53 General Unit Settings
Field Character Field Name Description
t CPU temperature calibration temperature °C
b Direct ADC value of CPU temperature diode 0 … 4096
l Heater control gain -100.0 ...[m] °C (default
-4.
0 °C)
m Heating set point °C
h Not used
Appendix E – Factory Configurations