EasyManua.ls Logo

Vaisala WXT530 Series - Appendix G. Complete Set of Accessories

Vaisala WXT530 Series
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 60 Wind Parameters
Parameter Factory Range Info Command
Wind data
con
trols
Order option
defined
1 = Dn Direction minimum Format:1111110001001000
The first 8 are for combined
message and the last 8 for
composite message.
1 = Data is on
0 = Data is o
Note that you write the
parame
ter format
1111110001001000 but the unit
answers 11111100&01001000.
WU,R
1 = Dm Direction average
1 = Dx Direction maximum
1 = Sn Speed minimum
1 = Sm Speed average
1 = Sx Speed maximum
0 = Reserved
0 = Reserved[& = Delimiter]
0 = Dn Direction minimum
1 = Dm Direction average
0 = Dx Direction maximum
0 = Sn Speed minimum
1 = Sm Speed average
0 = Sx Speed maximum
0 = Reserved
0 = Reserved
Wind update
in
t
erval
5 1 … 3600 seconds
WU,I
Wind
av
eraging
time
3 1 … 3600 seconds
WU,A
Wind speed
calcula
tion
mode
1 1 = Max/min calculation 3 =
Gust/lull calculation
WU,G
Wind speed
unit
M M = m/s, K = km/h, S = mph,
N = knots
WU,U
Wind
direc
tion
oset
0 -180 ... 180
WU,D
NMEA format W T = XDR, W = MWV
WU,N
Sampling rate 4 1, 2, 4 Hz
WU,F