Vaisala WXT530 Series - Send Identification Command (aI)

Vaisala WXT530 Series
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command format:
aR5<cr><lf>
a
Device address
R5
Supervisor message query command
<cr><lf>
Command terminator
Example of the response (the parameter set is configurable):
0R5,Th=25.9C,Vh=12.0N,Vs=15.2V,Vr=3.475V,Id=HEL___<cr><lf>
a
Device address
R5
Supervisor message query command
Th
Heating temperature (C = °C)
Vh
Heating voltage (N = heating is o)
Vs
Supply v
oltage (V = V)
Vr
3.5 V reference voltage (V = V)
<cr><lf>
Response terminator
Id
Information field
The content of the parameter Id is a te
xt string which you can modify with the Vaisala
Configuration Tool. The field can include customer-specific, additional information. For more
information on changing the settings, see the Vaisala Configuration Tool online help for the
Info field in the Device Settings window.
Mor
e Information
Supervisor Message (page 132)
7.2.8 Combined Data Message (aR)
This command requests all individual messages aR1, aR2, aR3, and aR5 with one
command.
C
ommand format:
aR<cr><lf>
WXT530 Series User Guide M211840EN-D