3 Prepare the e2695 Separations Module for Operation
100 July 29, 2013, 715003794 Rev. B
3. Press the Detectors screen key.
Result: A list of active devices and their IEEE-488 interface bus
addresses appears:
a. Press the Scan screen key to update the list.
b. Press the OK screen key to return to the Configuration screen, and
then press Exit.
Result: The separations module is ready to control Waters detectors
using separation methods and sample sets.
Controlled by Empower mode
In this remote control mode, Empower software controls the operation of your
HPLC system using the Ethernet interface or the IEEE-488 interface bus (but
not both simultaneously). Connect the Ethernet or IEEE-488 cable as
described in “Digital signal connections” on page 75.
To set the separations module to the Controlled by Empower mode:
1. Select the System field in the Configuration screen, and then press
Enter to display the list of operating modes.
2. From the list of operating modes, select Controlled by Empower, and
then press Enter.
Result: The network address field is highlighted.
3. Press Enter to display the network address drop-down list.
4. Depending on which interface you are using, make one of these
selections:
For Ethernet, select the Auto/Ethernet setting.
For IEEE-488, select an IEEE-488 address that is unused by other
chromatographic components connected to the Empower system.
5. Press Enter, and then press Exit.
Result: The separations module is ready to be controlled from the
Empower system.
Controlled by MassLynx mode
In this remote control mode, MassLynx software (version 3.5 or later) controls
the separations module (in 2690-emulation mode). MassLynx software is used